Technical characteristics
54 100,000 cts digital multimeters
Technical characteristics of the MTX 3292, MTX 3293 (continued)
In this setting, the user can measure a resistance.
Particular reference conditions:
The (+COM) input must not have been overloaded following the accidental application of a
voltage to the input terminals with the switch set to Ω or T°.
If this happens, the return to normal may take about ten minutes.
Protection: 1,414 Vpk
For measurements
above 5M
, a shielded
lead is recommended.
For a 2-wire link, use
very short wires
(<25cm) and twist them
together.
Automatic or manual range selection
“Active” protection by PTC thermistor
Measurement voltage: approx. 1.2V
Maximum open-circuit voltage: 3.5V typ.
In the 100M
range, in order to avoid the influence of the mains and guarantee the
stated specifications, it is best to disconnect the multimeter from the Wall Plug.
In this setting, the user can measure the capacitance of a capacitor.
Specified
measurement
range
For measurements
above 10nF, a shielded
lead is recommended.
For a 2-wire link, use
very short wires
(<25cm) and twist them
together.
() Use the REL function for values <10% of the range in order to restore the residual zero
(compensation for the capacitance of the leads)
Resolution 1,000 points
Automatic or manual range selection
“Active” protection by PTC thermistor
Maximum open-circuit voltage: 1V typ./4V max.