Technical characteristics
54 100,000 cts digital multimeters
Technical characteristics of the MTX 3292, MTX 3293 (continued)
Resistance
Ohmmeter
In this setting, the user can measure a resistance.
Particular reference conditions:
The (+COM) input must not have been overloaded following the accidental application of a
voltage to the input terminals with the switch set to Ω or T°.
If this happens, the return to normal may take about ten minutes.
Protection: 1,414 Vpk
Range
Accuracy
Resolution
Protection
1,000
0.1% L + 8 D
10m
1,414Vpk
10k
0.07% L + 8 D
100m
100k
1
1,000k
10
10M
1% L + 80 D
100
100M
3% L + 80 D R ≤ 50M
1k
For measurements
above 5M
, a shielded
lead is recommended.
For a 2-wire link, use
very short wires
(<25cm) and twist them
together.
Automatic or manual range selection
“Active” protection by PTC thermistor
Measurement voltage: approx. 1.2V
Maximum open-circuit voltage: 3.5V typ.
In the 100M
range, in order to avoid the influence of the mains and guarantee the
stated specifications, it is best to disconnect the multimeter from the Wall Plug.
100
Measure
Range
Accuracy
Resolution
Protection
100
0,2 % L + 10 D
0,01
1414 Vpk
Capacity
Capacitance meter
In this setting, the user can measure the capacitance of a capacitor.
Range
Operating
range
Specified
measurement
range
Resolution
Intrinsic error
Measurement
current
Measurement
time
1nF
0 to 1.000nF
0.100 to
1.000nF
1pF
2.5% L ± 15D
<10µA
≈400ms
10nF
0 to 10nF
0.1 to 10.00nF
10 pF
1% L ± 8D
<10µA
≈400ms
100nF
0 to 100.0nF
1 to 100.0nF
0.1nF
1% L ± 8D
<50µA
≈400ms
1,000nF
0 to 1,000nF
10 to 1,000nF
1nF
1% L ± 10D
<200µA
≈0.125s/µF
10µF
0 to 10.00µF
1 to 10.00µF
0.01µF
1% L ± 10D
<200µA
≈0.125s/µF
100µF
0 to 100,0µF
1 to 100.0µF
0.1µF
1% L ±10D
<500µA
≈0.125s/µF
1mF
0 to 1.000mF
0.1 to1.000mF
1µF
1% L ± 15D
<500µA
≈17s/mF
10mF
0 to 10.00mF
0.5 to10.00mF
10µF
1.5% L ± 15D
<500µA
≈17s/mF
For measurements
above 10nF, a shielded
lead is recommended.
For a 2-wire link, use
very short wires
(<25cm) and twist them
together.
() Use the REL function for values <10% of the range in order to restore the residual zero
(compensation for the capacitance of the leads)
Resolution 1,000 points
Automatic or manual range selection
“Active” protection by PTC thermistor
Maximum open-circuit voltage: 1V typ./4V max.