5.3 Mounting the electrode holder
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
916 TiTouch
5.3 Mounting the electrode holder
An electrode or, depending on the product version, a propeller stirrer can
be mounted onto the 916 TiTouch with the aid of the electrode holder.
NOTICE
To prevent damage to the electrode or rod stirrer, the permissible
immersion depth of the products applied can be fixed with the aid of
the clamping ring on the electrode holder.
Proceed as follows:
1
Push the clamping ring (6.2013.010) down as far as possible onto
the mounted support rod (6.2016.050). Make sure while doing so
that the indent is facing upwards.
2
Attach the electrode holder (6.2021.020) on the support rod.
3
Keep the locking lever pressed down and push the electrode holder
down.
4
Release the locking lever at the desired position. The electrode holder
is fixed.
5
Readjust the clamping ring to the electrode holder position and
screw tight to fix the position of the electrode holder. Make sure that
the nose of the electrode holder is placed in the opening of the
clamping ring.

Other manuals for Metrohm 916 Ti-Touch

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 916 Ti-Touch and is the answer not in the manual?

Metrohm 916 Ti-Touch Specifications

General IconGeneral
BrandMetrohm
Model916 Ti-Touch
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals