Sonicator Plus 930 Instruction Manual— Rev.J_09/13/17
39
Section 6—Maintenance and
Troubleshooting
6.1 Cleaning the Sonicator Plus 930
1. The Sonicator Plus 930 can be wiped off with a damp cloth. The power cord should be
disconnected from the unit before this is done. In the case of stubborn dirt a gentle
household cleaner can be sprayed on the cloth and then wiped on the unit. If this
method is used, remove any cleaner residue with a damp cloth. Do not spray cleaner
into the vents of the unit.
2. Follow the V Trode package insert for the use and care of the electrodes supplied with
the Sonicator Plus 930.
3. For routine cleaning of the electrode cables use soap and water. Thoroughly dry after
cleaning.
4. Use soap and water for routine cleaning of the Sonicator Plus 930 applicator. When
disinfection is necessary, use a disinfectant such as a 10% bleach solution. Rinse the
applicator thoroughly after disinfection to remove any residue. The Sonicator Plus 930
applicator is neither autoclavable nor gas sterilizable.
6.2 Routine Maintenance
1. Standard medical electrical safety checks should be performed annually by qualified
biomedical engineers or technicians trained to perform these procedures.
2. Inspect electrode cables and associated connectors for damage.
3. To assure accurate performance of the Sonicator Plus 930, calibration verification of
ultrasonic output should be performed on an annual basis.
4. Inspect treatment head for cracks, since they may allow ingress of conductive fluid(s).
5. Inspect treatment head cables and associated connectors for damage.
6. Avoid rough handling of the treatment head, since it is critical to the safe and effective
application of therapeutic ultrasound and relatively fragile.
6.3 Troubleshooting the Sonicator Plus 930
Symptom Action
1. Nothing lights when main
power switch is turned on.
Is line cord connected to outlet?
Does the outlet have power?
Unit may require servicing if none of the above
resolve the problem.