Section 9—Specifications
9.1 General Specifications:
Input: 90–240 VAC, 50–60 Hz, 2.3 Amp Nom.
Certification: The Sonicator Plus 930 complies with the ultrasound
performance standards set forth in the Code of Federal
Regulations, Title 21 (Food and Drugs), Part 1050.10
Weight: 8.7 pounds
Dimensions: 6 in (H) x 12 in (W) x 12 in (D)
Operating Temperature: +50°F to +104°F
Humidity:
Operating, 30% to 75% Relative Humidity at 104F
Nonoperating, 5 to 95% Relative Humidity, non-
condensing
Storage Temperature: -40°F to 167°F
Timer Accuracy: ±0.5 minutes for times less than 5 minutes
±10% for times from 5 to 10 minutes
±1.0 minute for times greater that 10 minutes
Maximum Treatment Time: 60 minutes–electrical stimulation
30 minutes–ultrasound or combination therapy
Treatment Timer: Treatment time counts down to zero when a time is set,
or up to 60 or 30 minutes when no time is set. The
digital timer indicates time in minutes and seconds. The
timer also indicates the remaining or elapsed treatment
time during the “Hold” period.
9.2 Ultrasonic Generator Specifications:
Frequency: 1.0 MHz ±10%
3.2 MHz ±10%
Modes: Continuous
Pulsed—20% duty cycle
Pulse Repetition Rate: 100 Hz ±5%
Pulse Duration: 2 msec ±5%, 20% duty cycle
Temporal Peak/ average
intensity ratio:
5:1 ±5%, 20% duty cycle
Maximum output power: 11 W
Maximum intensity: 2.2 W/cm²
Indication accuracy: ±20% (for any level above 10% of maximum)
Output description: The output waveform is continuous or pulsed as

Table of Contents