57
<Command> :MEASure:CURRent:DC <current NR3/disc>
:MEASure:CURRent:DC?
Function:It sets or queries the DC measurement current (LEVEL).
Description:
Set parameterThe value of DC currentis {0.0004~0.04|
MAXimum|MINimum}.
Set syntax
:MEASure:CURRent:DC 0.01
:MEASure:CURRent:DC 20m
:MEASure:CURRent:DC 1E-2
:MEASure:CURRent:DC MAXimum
:MEASure:CURRent:DC MINimum
Query syntax
:MEASure:CURRent:DC?
Return data 2.000000E-04 (Format is in <NR3>)
When the Level setup is voltage mode, the data returns 9.9E37.
<Command> :MEASure:ALC {OFF|ON|0|1}
:MEASure:ALC?
Function:It sets or queries the ALC.
Description:
Set parameter
ON|1, OFF|0
Set syntax
:
MEASure:ALC ON
Query syntax:MEASure:ALC?
Return dataWhen the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
<Command> :MEASure:SMONitor {0|1|OFF |ON }
:MEASure:SMONitor?
Function:It sets or queries the DISPLAY Vm/Im
Description:
Set parameter
ON|1, OFF|0
Set syntax
:
MEASure:SMONitor ON
Query syntax
:MEASure: SMONitor?
Return dataWhen the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
<Command> :FETCH:SMONitor:DC?
Function:It queries the DC voltage and current.
Description:
Query syntax
:FETCH:SMONitor:DC?
Return data 1.000000E-03,3.791975E-02 Vm,Im(Format is in <NR3>)
<Command> :FETCH:SMONitor:AC?
Function:It queries the AC voltage and current.
Description:
Query syntax
:FETCH:SMONitor:AC?
Return data 1.000000E-03,3.791975E-02 Vm,Im (Format is in <NR3>)
<Command> :MEASure:AVERage <averageNR1>
:MEASure:AVERage?
Function:It sets or queries the AVERAGE.
Description:
Set parameterThe value of the Average is {0~64}
Set syntax
:MEASure:AVERageage 10
Query syntax
: MEASure:AVERageage?
Return data 10 (Format is in <NR1>)