59
<Command> :MEASure:RANGe:DC {1|2|3|4|AUTO|HOLD}
:MEASure:RANGe:DC?
FunctionIt sets or queries the DC RANGE.
Description
Set parameter{1|2|3|4|AUTO|HOLD}
Set syntax
:MEASure:RANGe:DC 2
:MEASure:RANGe:DC AUTO
Query syntax
:MEASure:RANGe:DC
Return data 2 (Format is in <NR1>)
The range recommends setting as<Auto>in order to obtain better measuring accuracy. The
range actually measured will be displayed at the lower-left corner of LCD panel. Further, there
are also <1-30Ω all frequency>, <2-300Ω frequency below 1.2 MHz>, <3-3KΩ frequency be-
low 120 KHz> and<4-30KΩ frequency below 12 KHz> ranges for option. Faster measuring
speed can be achieved when setting at “Range Hold”.
<Command> :MEASure:RANGe:AC {1|2|3|4|AUTO|HOLD}
FunctionIt sets or queries the AC RANGE.
Description
Set parameter{1|2|3|4|AUTO|HOLD}
Set syntax
:MEASure:RANGe:AC 3
:MEASure:RANGe:AC AUTO
Query syntax
:MEASure:RANGe:AC
Return data 3 (Format is in <NR1>)
The range recommends setting as<Auto>in order to obtain better measuring accuracy. The
range actually measured will be displayed at the lower-left corner of LCD panel. Further, there
are also <1-30Ω all frequency>, <2-300Ω frequency below 1.2 MHz>, <3-3KΩ frequency below
120 KHz> and<4-30KΩ frequency below 12 KHz> ranges for option. Faster measuring speed
can be achieved when setting at “Range Hold”.
<Command> :MEASure:OIMPedance {100|25}
:MEASure:OIMPedance?
FunctionIt sets or queries the RO.
Description
Set syntax
:
MEASure:OIMPedance 100
Query syntax
:MEASure:OIMPedance?
Return
data 100(Format is in <NR1>)
<Command> :MEASure:COMParator:PARAmeter {1|2|3|4}
:MEASure:COMParator:PARAmeter?
FunctionIt sets or queries which parameters number is to be compared.
Description
Set syntax
:MEASure:COMParator:PARAmeter 3 (After this command, every compared setup
is at this parameters number.)
Query syntax
:MEASure:COMParator:PARAmeter?
Return
data3 (Format is in <NR1>)
<Command> :MEASure:COMParator:STATe {OFF|ON|0|1}
:MEASure:COMParator:STATe?
FunctionIt sets or queries the COMP open or not.
Description
Set syntax
:
MEASure:COMParator:STATe ON
Query syntax
: MEASure:COMParator:STATe?
Return
dataWhen the setup is OFF, it returns 0.
When the setup is ON, it returns 1.