73
<Command>
:LIST:RETest {OFF|STEP|ALL|0|1|2}
:LIST:RETest?
Function:It sets or queries the FAIL RETEST of list mode.
Description:
Set parameterIt retests continuously until the test result is pass, when the test result is fail. If
selecting <OFF>, it doesn’t execute this function. If selecting <STEP 1>, only the first step is
going to retest. If selecting <ALL>, no matter which step occur fail result, the fail step is going
to retest until the result become pass.
Set syntax
:LIST:RETest STEP
:LIST:RETest 2
:LIST:RETest OFF
Query syntax
:LIST:RETest?
Return dataOFF | STEP |ALL (Format is in <disc>)
<Command>
:LIST:FILE:LOAD <filename>
Function:It opens the list mode's file.
Description:
Set syntax:MEASure:FILE:LOAD MICROTEST.(open the MICROTEST list mode file)
<Command>
:LIST:FILE:LOAD?
Function:It queries the list mode's filename which is being using.
Description:
Query syntax:MEASure:FILE:LOAD?
Return data MICROTEST
<Command> : LIST:BIAS:VOLTage <NR3>
: LIST:BIAS:VOLTage?
Set parameter: The value of BIAS Voltage -12.000~12.000
Function:It sets or qerties the BIAS Voltage of list mode.
Description:
Set syntax: : LIST:BIAS:VOLTage 6
Query syntax: LIST:BIAS:VOLTage
?
Return
data +6.000000E+00
<Command> :LIST:FILE?
Function:
It queries all of the list mode's filenames that are stored in the memory.
Description:
Query syntax: :LIST:FILE?
Return
data
N,filename1,filename2,filename3,...,filenameN
where N= total file count
<Command> :LIST:STATistic {ON|OFF|0|1}
:LIST:STATistic?
Function:
It sets or queries the statistic function in the list mode.
Description:
Set parameter {ON|OFF|0|1}
Set syntax: LIST:STATistic 1
Query syntax: LIST:STATistic?
Return
data
0|1
where 0= statistic function is disabled.
1= statistic function is enabled.