2"
(5 cm)
Clearance
21"
(53 cm)
Support Surface
1"
(3 cm)
5"
(13 cm)
Requirements
5"
(13 cm)
18"
(46 cm)
M11
M9/M9D
M11
23" (58 cm)
M9/M9D
22" (56)
2"
(5 cm)
Overhang/Shelf
M9/M9D
18"
(46 cm)
M11
15"
(38 cm)
Installation
Operating Environment
Ambient Temperature Range: 68°F to 104°F (+20°C to +40°C)
Relative Humidity: < 80% (non-condensing)
(Pollution Degree 2, in accordance to IEC664)
Normal Operating Altitude: < 9842 ft. (3000 m) above sea level
Device approved for INDOOR USE ONLY.
Device to be operated in a relatively dust-free environment.
(Pollution Degree 2, in accordance to IEC664)
Device should be connected to a power source with over-voltage limits less than 1500 watts from
mains to ground. (Installation Category II in accordance to IEC664)
The M9 / M9D and M11 will emit 5000 BTU / HR during operation.