Weekly Maintenance
Tray Rack
Tray Plate
Trays
Clean Chamber / Trays (including Rack and Plate)
Equipment Alert
Failure to change water may result in sterilizer malfunction. Do not use bleaching
agents or any abrasive materials / substances in chamber (i.e. bleach, steel wool, wire
brush, scouring powder, etc.). Failure to comply may result in damage to the chamber and/or
other components.
Note
A) Disconnect the upper portion of the reservoir drain tube from the panel clips, bend it downward,
and drain the reservoir water into a suitable container, e.g. a bucket, and dispose of the water.
B) Remove the trays, tray rack, and tray plate from the sterilizer.
(Refer to the following page for instructions on removing / installing the tray rack and tray plate).
C) Wash trays, rack, plate, and inside of chamber with mild soap or Speed-Clean and distilled water or
water that meets the referenced water purity specifications.
D) Refill reservoir with distilled water or water that meets the referenced water purity specifications.
Caution
To prevent burns, allow unit to cool before
draining reservoir.