Operation
Before Operating the Sterilizer
Filling the Reservoir
Caution
Programmable cycles 1 and 2 are provided for those applications requiring sterilization
parameters different than the preset cycles. These cycles are not FDA cleared for medical
use. All material processed in these cycles must be validated by the user to ensure sterility of the
processed load.
WARNING
Do not use this sterilizer for sterilizing volatile substances or for any purpose other
than its intended design. Burns and toxic or explosive conditions could result.
Do not force door handle at any time. Chamber pressure may cause door to open with
extreme force. If door handle does not move freely, allow unit to cool and depressurize for 40
minutes before opening door. Failure to adhere could result in serious personal injury.
Do not run the sterilizer without the tray plate in place. If the sterilizer malfunctions,
immediately unplug sterilizer, and call for service; do not attempt to repair the sterilizer
yourself. Doing so could result in serious injury.
Equipment Alert
For optimal performance, allow the sterilizer to reach room temperature before operating.
To fill reservoir...
A) Open door to unit.
B) Pour distilled water into fill opening until
water level reaches the top of the fill level
label on the water level indicator tube.
Equipment Alert
Use distilled water or water that meets the
referenced water purity specifications.
Failure to comply may result in sterilizer
malfunction and/or premature failure due
to excessive corrosion.