English - 29
003-2915-99
© Midmark Corporation 2018
Programmable Cycle Buttons
Cycle buttons can be used for custom applications that are not covered by standard cycle
user to adjust cycle parameters in order to sterilize items that cannot be sterilized in any of the standard
or button. Use the instructions in the chart below to set time / temperature parameters for these
buttons. (If you wish to change the settings, these buttons may be reprogrammed at any time).
NOTE:
Action Description Display
Press desired
button
This selects the button that will be
programmed.
Press This brings up the sterilization
temperature programing display.
Caution
Sterilization temperature can be adjusted from a minimum of 250°F (121°C) to
a maximum of 275°F (135°C). At high altitudes sterilization temperatures above
270° F (132° C) may cause the pressure relief valve to leak.
Adjust Sterilization
Temperature
The “+” and “-” buttons adjust the
temperature by 1°increments.
When the desired temperature appears
on the display...
This continues on the following page...
Note:
The programmable cycles can
be programmed for the monthly
cleaning cycles with no dry time
to expedite monthly maintenance.