Programmable Cycle Buttons
Cycle buttons can be used for custom applications that are not covered by standard cycle

user to adjust cycle parameters in order to sterilize items that cannot be sterilized in any of the standard

or button. Use the instructions in the chart below to set time / temperature parameters for these
buttons. (If you wish to change the settings, these buttons may be reprogrammed at any time).
NOTE:

Action Description Display
Press desired
button
This selects the button that will be
programmed.
Press This brings up the sterilization
temperature programing display.
Caution
Sterilization temperature can be adjusted from a minimum of 250°F (121°C) to
a maximum of 275°F (135°C). At high altitudes sterilization temperatures above
270° F (132° C) may cause the pressure relief valve to leak.
Adjust Sterilization
Temperature
The “+” and “-” buttons adjust the
temperature by 1°increments.
When the desired temperature appears
on the display...
This continues on the following page...
Note:
The programmable cycles can
be programmed for the monthly
cleaning cycles with no dry time
to expedite monthly maintenance.