English - 30
003-2915-99
© Midmark Corporation 2018
NOTE:
Action Description Display
Press To store the temperature. This brings up
the sterilization time programing display
showing the current value.
Caution
Sterilization time can be adjusted from a minimum of 3 minutes to a maximum
of 45 minutes. Because it is important to properly coordinate the cycle time with
the sterilization temperature the minimum selectable time will be determined by
the temperature setting.
Adjust Sterilization
Time
The “+” and “-” buttons adjust the
time by 1 minute increments.
When the desired time appears on the
display...
Press To store the sterilization time. This brings
up the vent speed programing display and
the current value.
Adjust Vent Speed
vents the chamber.
fraction of a second (once per minute)
to slowly vent the chamber.
When the desired setting appears on the
display...
Press To store the setting and bring up the dry
time programing display and the current
value.
This chart continues on the following page...
Programmable Cycle Buttons - continued...