Location Requirements - continued...
Support Surface


 

Dimensions
Depth (front to back) M11 - 21” (53 cm) M9 / M9D - 18” (46 cm)
Clearance Requirements



the surrounding area.
Back of Unit - Back Wall ................................. 2” (5 cm)
Front Support Surface - Front Sterilizer ........ 1” (3 cm)
Sides of Unit - Side Wall ................................ 2” (5 cm)
 .......... 5” (13 cm)
 ..... M11 - 18” (46 cm) M9 / M9D - 15” (38 cm)
Under Cabinet or Shelf................................... M11 - 23” (58 cm) M9 / M9D - 22” (56 cm)
Relocation Requirements for Sterilizer
Disconnect power cord from electrical outlet and allow sterilizer to cool.

WARNING
For 115 VAC models: Use 104 - 127 VAC, 50/60 Hz alternating current only.
For 230 VAC models: Use 207 - 253 VAC, 50/60 Hz alternating current only.
Failure to do so may result in electric shock to personnel and / or damage to sterilizer.
Note
For safety, the unit must be connected to a properly polarized and grounded receptacle. Always use a
power cord with grounding connections that match the receptacles in your location.
115 VAC Unit: 


230 VAC Unit: 


* Power source must have over-voltage limits less than 1500 watts from mains to ground.
(Installation Category II in accordance to IEC664)