Informational Messages
The chart below lists the informational messages that may appear during operation.
Message Possible Cause Solution
INITIALIZING SYSTEM
Unit power cord was just plugged in -
standard informational
message.
Normal operation will occur after a
4 second pause.
TOTAL CYCLES XXX
M9, VX.X.X
Unit power cord was just plugged
in - standard informational message.
(M9 / M9D or M11) and software
version)
Normal operation will occur after this
message.
MM/DD/YYYY 8:07 Am
VXXXXXXX
Unit power cord was just plugged
in - standard informational message.
Normal operation will occur after this
message.
PERFORM WEEKLY
MAINTENANCE
and 21 days after the unit is plugged
into a power source to prompt the
operator to perform weekly maintenance
described in this manual.
Perform weekly maintenance.
The message will automatically clear
PERFORM MONTHLY
MAINTENANCE
This message is displayed every 28
days to prompt operator to perform
monthly maintenance described in this
manual.
Perform monthly maintenance.
The message will automatically clear
Caution
If an error occurs more than once, stop using the sterilizer. Record the message or error
code, unplug unit, then call an authorized service representative. (see “Calling for Service”)
If an error message includes: “Items Not Sterile”, the items in sterilizer shall not be considered sterile;
they must be run through a successful sterilization cycle.
Action Description Display
Press and hold
during power up
codes displayed on the unit.
NOTE: “1” is the most recent
error code; “5” is the oldest.
Press the “stop” button
the stop button.
Accessing the last 5 Error Codes
Error Messages