Error Messages - continued...
Message Possible Cause Solution
C010: POWER UP MODE
SYSTEM PWR LOSS
ITEMS NOT STERILE
PUSH STOP TO RESTART
Unit had loss of power
during cycle.
Press STOP button to restart.
CO60: POWER UP MODE
SYSTEM HARDWARE
Power was interrupted
glitch.
Unplug unit power cord for 1 minute and then plug
service representative. (see “Calling For Service”)
C102: FILL MODE
STOP PRESSED
STOP button was pressed
during cycle.
Press STOP button to restart.
C103: THROUGH C105
HEATUP, STERILIZE OR
VENT MODE
STOP PRESSED
STOP button was pressed
during cycle.
/ temperature dissipates. Press STOP button to return
to select cycle mode where a new cycle may be
initiated.
C106: DOOR MODE
STOP PRESSED
STOP button was pressed
during cycle.
Press STOP button to restart.
C232: FILL MODE
WATER LOW
Water level in
reservoir is too low.
return to select cycle mode where a new cycle may be
initiated.
of chamber) is clogged.
(see “Monthly Maintenance”)
C326: DOOR MODE DOOR
CLOSED
Door latch safety switch is
still making contact after
door motor operated.
Open door.
C382: FILL
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Close the sterilizer door.
(Cycle will continue where left off.)
C383: HEATUP MODE
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
-
perature dissipates. Press STOP button to return to
C384: STERILIZE MODE
DOOR OPEN
Sterilizer detects that door
safety switch contacts
opened.
Unplug unit power cord for 1 minute and then plug
service representative. (see “Calling For Service”)
C533 through C633:
STEAM TEMP LOW
or
STEAM TEMP HARDWARE
or
PRESSURE LOW
Sterilizer detects that
temperature and / or
pressure is outside the
limits for normal
operation.
Unplug unit power cord for 1 minute and then plug
service representative. (see “Calling For Service”)
This chart continues on the following page...