English - 39
003-2915-99
© Midmark Corporation 2018
Example of Typical Printout of a Program Cycle
Midmark M9 - v0.1.1
Total Cycles: 10
Operator
Temp: 270 Degrees F
Time: 3 Minutes
Dry: 30 Minutes
mm:ss Degrees
0:00 77.2F 0.0
2:00 82.7F 0.7
4:00 152.2F 5.4
mm:ss Degrees
0:00 272.1 F 28.0
1:00 273.0 F 28.7
2:00 272.2 F 27.9
3:00 272.4 F 28.7
Min 272.0 F 27.9
273.1 F 29.0
Drying Start
mm:ss Degrees
5:00 192.8 F 0.0
10:00 192.7 F 0.0
30:00 192.8 F 0.0
Drying Complete
0:36
15:40
3:00
2:28
30:00
00:51:44
--------------------------------------------
Sterilizer Identication
(Serial Number)
Total # of cycles
run on unit
Line provided for
operator signature
Date and Time
of cycle start
Selected Cycle
Indicates Filling
phase initiated
Sterilization temperature,
time, and dry time settings
During Heat-Up phase, the
printer records chamber
temperature and pressure
in 2 minute increments
During Sterilization phase,
the printer records chamber
temperature and pressure in
1 minute increments
Summary of
Sterilization phase
Indicates Venting
phase initiated
During Drying phase,
the printer records time in
5-minute increments
These lines will show the
total dry time completed
Cycle summary
Model # and
Software Version
These lines will not print
if Dry Cycle is aborted
before it completes