English - 8
003-2915-99
© Midmark Corporation 2018
Controls and Indicators

The accompanying tables describe the function of each control / indicator.
Control Function
Display See illustration 


remaining cycle time is displayed as well as temperature and
pressure. Display also shows error message if a malfunction
Troubleshooting section of this manual for a
 Informational / Error Messages.
Unwrapped button

270°F (132°C) for 3:00 minutes / 30 minute drying cycle.
Pouches button

paper / plastic sterilization pouches or wrapped instruments
at: 270°F (132°C) for 4 minutes / 30 minute drying cycle.
Packs button

250°F (121°C) for 30 minutes / 30 minute drying
cycle.
Handpieces
button

270°F (132°C) for 4 minutes / 30 minute drying cycle.
Start button

pressing Start will activate heater for 10 minutes.
Stop button
Terminates selected program or function.
1 or 2 buttons

run in these cycles
must be validated
for sterilization by
the user.
Programmable cycle buttons that allow an operator to create two
different programmed cycles for special applications. Sterilization
-
dure can be adjusted or changed.
Caution
These cycles are not FDA cleared and validation of sterility
of items processed using them is the responsibility of the
user.