EasyManua.ls Logo

Midmark M9 Series - Preparing Items for Sterilization; Immediate Use Sterilization

Default Icon
156 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English - 17
003-2915-99
© Midmark Corporation 2018
Preparing Items for Sterilization


manufacturer’s instructions for proper cleaning and preparation of the device for sterilization should always
be followed:


recommended over manual cleaning for clinician safety and cleaning effectiveness.




packaging.


Immediate Use Sterilization
The M9 / M9D and M11 are capable of Immediate Use sterilization - sterilizing unwrapped instruments for


another (See Photo 1).
WARNING
Clean and dry instruments thoroughly before placing them into tray. Improper cleaning may
result in non-sterile instruments or damage to the unit. Follow instrument manufacturer’s
guidelines and CDC recommendations for handling and cleaning instruments prior to sterilization.
Equipment Alert
Do not sterilize items composed of any of the following materials in the M9 / M9D or M11.
• Corrosion sensitive metal (ex. carbon steel, iron, etc.)
• Fragile items susceptible to breaking under pressure / high temperature
• Biomedical waste
• Plastics that may break down or produce residue when exposed to steam / high
temperatures.
Examples


Types of Items that can be processed in the M9 / M9D and M11 - Continued...

Table of Contents

Related product manuals