English - 23
003-2915-99
© Midmark Corporation 2018
M11 Maximum Capacities **
Load Type Large Tray Small Tray Total
or or or
Handpieces 2 handpieces with other
instruments
2.7 lbs. (1225 grams) or
2 handpieces with other
instruments
1.8 lbs. (816 grams) or
8 handpieces with other
instruments
9.0 lbs. (4082 grams) or
Packs* 145 in
3
(2376 cm
3
108 in
3
(1770 cm
3
505 in
3
(8275 cm
3
)
* Allow a minimum of 1/4” (6.4 mm) space between each pack, and from the chamber wall.
** The default dry time may need increased due to variations in load conguration, wrapping materials, and the environment to
completely dry the chamber contents at these capacities.
Load Size
Successful sterilization is dependent on correct loading of the sterilizer. Do not overload the sterilizer
steam circulation and adequate drying. The charts below are provided as a reference regarding the
processed to assure good steam circulation and drying should be the guiding factor in determining the
M9 / M9D Maximum Capacities **
Load Type Large Tray Small Tray Total
2.4 lbs. (1089 grams) or 1.6 lbs. (726 grams) or 8.0 lbs. (3629 grams) or
Handpieces 2 handpieces with other
instruments
2.4 lbs. (1089 grams) or
2 handpieces with other
instruments
1.6 lbs. (726 grams) or
8 handpieces with other
instruments
8.0 lbs. (3629 grams) or
Packs* 90 in
3
(1475 cm
3
55 in
3
(901 cm
3
290 in
3
(4752 cm
3
)
Equipment Alert
Do not use towels or packaging containing chlorine bleach
residue. Failure to comply may cause rusting or discoloration
sterilizer.
Textile Loads
proper steam circulation and drying. (Photo 15)
Photo 15