English - 36
003-2915-99
© Midmark Corporation 2018
Remove / Clean Door and Dam Gaskets
A) Remove door and dam gaskets from chamber door,
then remove the gasket ring from the door gasket.
B) Clean gaskets and ring with Speed-Clean,
distilled water, and a soft brush.
C) Inspect gaskets for damage / shrinking / swelling.
Replace gaskets if damage is apparent.
D) Press gasket ring into the channel in the
door gasket and reinstall the gasket in the door.
E) Install dam gasket.
Equipment Alert
Use care when wiping the inside of the chamber. Failure to comply may result in
damage to the heating element, steam temperature probe, and/or level sensor probe.
Equipment Alert
Do not operate sterilizer without filters in place.
H) Remove and clean filters. The filters are intended to prevent debris from causing valve failures.
Between regular monthly cleanings if the fill or vent times become too long or items will not dry the
filters should be cleaned. (Refer to the illustration for location of filter screens.)
I) Grasp filter and gently pull away from chamber wall while twisting slightly.
(If necessary, pliers may be used to remove filters)
J) Clean filters with Speed-Clean and distilled water. A small stiff bristled brush or ultrasonic cleaner
may be helpful. Rinse filters with distilled water. Replace filter(s) if debris cannot be removed by
cleaning.
K) Wipe out the inside of chamber.
L) Install filters. (Press inward, toward chamber wall while twisting slightly).
M) Install tray plate, rack, and trays.
Clean Chamber / Plumbing - continued...