Technical Information
31
PW 6xx7
No. Program step Basic
setting
Setting options Notes
6 Temperature
Cold Cold Water intake temperature, control does not
activate any heating.
15°C to 95°C Selection of heating temperature for this
block from 15°C to 95°C in 1°C steps.
7 Hysteresis
Normal Normal Target temperature +0°C / -6°C in this
block.
Small Target temperature +3°C / - 0°C in this
block.
User-defined Values of upper and lower limits can be
modified in 1°C steps. Smallest permitted
difference between upper and lower limits
3°C. Standard settings: Hysteresis off:
+0°C and Hysteresis on: -6°C. Upper
range limits: 0°C to +3°C, lower range
limits -10°C to 0°C with smallest
permissible difference 3°C.
8 Warmup
Lock symbol
(if No has
been pro-
grammed
under
Heating in
this table/
block) or No
(if Yes has
been pro-
grammed
under
Heating in
this table/
block)
Lock symbol No has been programmed under Heating
in this table/ block.
No Do not use any programmed warmup rate.
1°C/min to 10°C/min Warmup rate can be modified between
1°C/min and 10°C/min in 1°C steps in this
block.
9 Level 1
(dependent on
the setting for
Fill type in the
program header
for this program)
Auto load
adjustment: 0
mm
0 to 350 mm Level 1 in mm wc (water column) can be
modified in 2 mm wc steps. The lower limit
is the minimum level for heating.
Water height
in mm: 0 mm
0 to 350 mm Level 1 in mm wc (water column) can be
modified in 2 mm wc steps. The lower limit
is the minimum level for heating.
Water
quantity in L:
0L
0 to permitted upper limit Level 1 in L in 1L steps: PW 6107 max. 95
L, PW 6137 max. 120 L, PW 6167 max.
125 L, PW 6207 max. 152 L.
Load-to-
liquor ratio in
kg/L: 1: 0.0
kg/L
1:0.0 kg/L, 1:0.5 kg/L,
1:1.0 kg/L, ..., to 1:20.0
kg/L
Level 1 as laundry:suds ratio in 0.5 kg/L
steps.
10 Intake path 1
Automatic Automatic Depending on the program block and the
current water temperature in the machine,
the control selects which valves to
operate. In order to save energy the
control will attempt to achieve the desired
wash cycle temperature during the first
water intake, if possible, by operating the
cold and hot water valves alternately. For
details, see tables at Section 100-4.2.
User-defined: Valve 1,
valve 2, valve 3, valve 4
The programmer selects No for water inlet
valves 1 to 4 or, for each valve, 1 of 11
possible options from the list K1, K2, K3,
K4, K6, KD, KH, WD, W1, W2. For details,
see tables at Section 100-4.2.