CAUTION
If the pressure reaches 53 psi (3.6 bar) and the pump does not
automatically stop, press the control knob to abort the test. Refer to
Pressure Sensors Test Issues, Overpressure.
NOTE
If no pressure is detected after about 30 seconds, the test stops. See
Pressure Sensors Test Issues, No Pressure Increase Detected.
19. As soon as the pump stops, read the pressure on the pressure gauge. The
pressure should be between 37 psi (2.5 bar) and 53 psi (3.6 bar).
NOTE
If the pressure is not between 37 psi (2.5 bar) and 53 psi (3.6 bar), refer
to Pressure Sensors Test Issues, Pressure Outside the Acceptable
Range.
20. Press the
button. The following screen displays:
Remove the pressure control kit for Steritest
®
Pumps to release the pressure
in the canister.
Remove the bottle from the bottle holder basket and place it on the
workbench.
Open the clamp on the bottle side.
Classe:1.6
316 SS
TUBE AND CONNECTION
MILLIPORE
30
2
3
4
bar
psi
1
2
20
10
0
40
50
60
37 psi
53 psi