Page 96 of 113
Check the tubing and do one of the following:
If the tubing is damaged:
Turn and press the control knob to select Yes. The following screen
displays:
Discard the Steritest
®
device and unpack a new one. Press the
button.
Restart the pressure sensor test procedure as explained in Configuring
the Pump, Pressure Sensor Test.
If the tubing is not damaged,
Turn and press the control knob to select No. The following screen
displays:
Re-install the tubing in the pump head.
Press the
button. The screen for the selection of the pressure
sensor to verify displays with the last sensor tested selected.
Press the control knob to confirm and restart the procedure from
step 7 (step 5 on the pump screen) or step 15 (step 13 on the pump
screen) of the Pressure Sensors Test procedure.