Allow system to run for 5 minutes
to vessel.
Close point-of-use valve. Turn
to RECIRCULATION.
Allow system to recirculate
for 5 minutes.
Open point-of-use valve. Turn
to PRODUCTION.
Allow system to run to vessel
for 2 minutes.
Close point of use valve. Turn
to RECIRCULATION.
Press OPERATE/STANDBY to switch
to standby mode.
Remove adapter and tubing
from point-of-use.
Put 1-2 turns of Teflon tape on
threaded end of Millipak final
filter unit.
Screw Millipak firmly into
point-of-use adapter until
the vent cap is oriented to
the system.
Fit Millipak inside point-of-use
cover.
Open vent cap on Millipak.
Turn point-of-use valve to
PRODUCTION.
Threaded adapter
Vent cap
Press OPERATE/STANDBY to put
the system into operate mode.