8.4 Failed elution of labeled cells
In very rare cases the elution buffer does not run through the columns of the Multi-24 Column Block after the
vacuum impulse has been applied. If so,
l plug the column air-tight, for example with a glove covered thumb, to create a minimal overpressure that re-
engages elution buffer flow through.
l ensure that the plug is sterilized, if sterile conditions are required.
8.5 Poor elution recovery
If the elution recovery is poor, ensure that:
l base and lid of the MACS Elution Chamber are assembled correctly without tilting.
l the lid of the MACS Elution Chamber and labware sealing are dust free, intact, and positioned correctly.
If the problem persists or if a sealing has to be changed, please contact Miltenyi Biotec Technical Support.
8.6 Error codes
All errors are coded. If an error is shown, read the following table for the description of the error and potential
solution.
Error message Description and solution
STEPPING LOSS ERROR –101 INITIALIZING SYSTEM Motor out of sequence.
Tap OK for intermediate reinitialization of the system.
WRONG MAG POS ERROR –102 INITIALIZING SYSTEM Incorrect positioning of the MultiMACS Cell24 Magnet.
Tap OK for intermediate reinitialization of the system.
WRONG COL POS ERROR –106 INITIALIZING SYSTEM The MultiMACS Column Holder is locked in wrong
position.
Tap OK for intermediate reinitialization of the system.
The MultiMACS Column Holder will be repositioned.
POSITION FAILED ERROR –111 INITIALIZING SYSTEM Incorrect positioning of the MultiMACS Cell24 Magnet.
Make sure that the instrument is not exposed to direct
sunlight. Tap OK for intermediate reinitialization of the
system.
CANNOT PICKUP ERROR –117 INITIALIZING SYSTEM Locking of MultiMACS Column Holder did not work.
Tap OK for intermediate reinitialization of the system.
WRONG SIGNAL ERROR –118 CALL TECHNICAL
SERVICE
Electronic hardware problem.
Contact Miltenyi Biotec Technical Support.
LOCK FAILURE ERROR –123 CALL TECHNICAL
SERVICE
Electronic or mechanical hardware problem.
Contact Miltenyi Biotec Technical Support.
ELUTION STATION NOT DETECTED ERROR –401 The MACS Elution Station is not detected by the
MultiMACS Cell24 Separator.
Ensure that the cable of the MultiMACS Cell24 Separator
is connected correctly to the MACS Elution Station.
Table 8.1: Errorcodes of the MultiMACS Cell24Separator Plus.