2.1.2 Accessories
Product Order no.
MultiMACS 5 mL Tube Rack, 1 piece 130-095-331
MultiMACS 12× Single-Column Adapter LS, 1 piece 130-108-816
Table 2.2: MultiMACS Cell24Separator Plus accessories.
2.1.3 Consumables
Product Order no.
Multi-24 Column Block, 1 piece 130-095-692
Multi-24 Column Blocks, 8 pieces 130-095-691
LS Columns, 25 pieces 130-042-401
LD Columns, 25 pieces 130-042-901
Whole Blood Column Kit, 1 piece 130-093-545
24-well Deep Well Plates, 2x2 pieces 130-110-500
Single-well Deep Well Plates, 4 pieces 130-114-966
2-well Deep Well Plates, 4 pieces 130-120-010
3-well Deep Well Plates, 4 pieces 130-120-009
4-well Deep Well Plates, 4 pieces 130-120-008
Deep Well Plate Lids, 8 pieces 130-120-007
Table 2.3: MultiMACS Cell24Separator Plus consumables.
Use deep well plates that comply to ANSI/SBS standards (length 127.76 mm ± 0.5 mm, width 85.48 mm ±
0.5mm). As height of the deep well plates a value of 44.0 mm is set by default. The height value can be
changed manually.
2.2 Assembly for benchtop use
Leave a space of at least 20 cm distance above the MultiMACS Cell24 Separator Plus to ensure that the
instrument is able to tilt backwards, thus minimizing damage. If an object has been accidentally placed
under the down-moving MultiMACS Cell24 Magnet, the physical resistance of the object causes the
instrument to tilt backwards similar to a hydraulic jack.
1 Unpack all components.
2 Place the MultiMACS Cell24 Separator on a clean, stable, and flat laboratory bench. For sterile conditions
place the instrument in a laminar flow cabinet.
3 Lift the front side of the instrument and attach the Tip-Touch Plate.
2
INSTALLATION
44