4-
Parameter Description 
Reagent blan
absorbance  
Absorbance range of the reagent blank. Once this range is 
exceeded, error message will appear. 
The reagent blank absorbance should be entered from low to 
high. The high and low limit should not be equal. The default is 
void. 
 
NOTE 
The absorbance of reagent blank is entered in the 
format of real value×10000. 
 
Absorbance 
limit 
The maximum or minimum absorbance of the test result. For 
increase reaction, system will alarm when the reaction 
absorbance is higher than the absorbance limit; for decrease 
reaction, system ill alarm when the reaction absorbance is lower 
than the absorbance limit (for Kinetics and Fixed-time analysis, 
the absorbance limit is seen as the substrate depletion limit).     
 
NOTE 
The absorbance limit  is  entered in the format of 
real value×10000. 
 
Lin. Limit 
It is only valid for Kinetics. The range is 0-1. The default is void. 
0 means no check. 
By Diff.  Set combinations based on sample type and patient gender, and 
enter respective reference range for each combination. Those 
who are under the age of 12 are taken as children. 
 
 
The following table introduces the buttons on the page. 
Buttons  Function 
Save  Save the test information. 
Cancel  After selecting a test and setting or modifying parameters for it, click 
this button to cancel the setting or modification. 
Return  Click this button to return to the main screen. 
4.2.1.3 Calibration 
The  Cal Details screen is where you can set the test calibration information 
including calibration rule, calibrator concentration and replicates.