EasyManuals Logo

Mindray BA-88A Operation Manual

Mindray BA-88A
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
4-
7
Figure 4-5 Cal Details
The following table explains the parameters on the Cal Details screen.
Parameter Description
K factor For Absorbance analysis, the K factor is fixed at 1 and can not be
edited.
For Kinetics analysis the K factor is selected by default and you
can enter according to the reagent instruction. Decimal is allowed.
Rule Calibration rule.
Replicate It refers to the times of each calibration test. It ranges from 1-3. The
default is 1.
Set the number of calibrators and the calibrator set point. The set
point of the calibrator is arranged from lowest to highest. The void
means 0 concentration. If set points are not set according to the
above rules, error message will appear.
S1-S8
NOTE
After the information about the number of the
calibrators and the set points are modified, the
existing information about calibration parameters
and calibration curve will be deleted, so recalibration
is required.
The following table introduces the buttons on the screen.
Buttons Function
Save Save the test information.
Cancel After selecting a test and setting or modifying parameters for
it, click this button to cancel the setting or modification.
Return Click this button to return to the main screen.

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
Light sourceHalogen lamp
Display7 inch touch screen
PrinterThermal printer
InterfaceRS232
Wavelengths340, 405, 546, 578, 630nm
Carry-over< 1%
Power SupplyAC 100~240V, 50/60Hz
Operating Temperature15-30°C

Related product manuals