BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual          3-13 
3.3.4 NIBP Tests 
See section 1.2.21.2.2NIBP Tests. 
 
 
3.3.5 Temp Test 
Tools required: 
  Resistance box (with accuracy above 0.1 Ω)   
 
1.  Connect the two pins of any Temp connector of a module to the two ends of the resistance box using 
two wires. 
2.  Set the resistance box to 1354.9 Ω (corresponding temperature is 37ºC). 
3.  Verify each Temp channel of the monitor and make sure that the displayed value is within 37±0.1ºC. 
4.  Repeat steps 1 to 3 to verify each Temp channel of the monitor. 
 
 
3.3.6 IBP Tests 
Performance Test 
Tools required: 
  Patient simulator Medsim300B, MPS450, or other equivalent equipment 
  Dedicated IBP adapter cable (P/N 00-002199-00 for 300B, P/N 00-002198-00 for MPS450) 
 
1.  Connect the patient simulator to the monitor's IBP connector. 
2.  Set the patient simulator output to the IBP channel to 0 mmHg. 
3.  Press the Zero key on the module to make a zero calibration. 
4.  Set static pressure to 200 mmHg on the patient simulator. 
5.  The displayed value should be within 200±2 mmHg. 
6.  If the error is beyond ±2 mmHg, calibrate the IBP module. If the IBP module was calibrated with a 
dedicated reusable IBP sensor, check the calibration together with this IBP sensor. 
7.  Make the patient simulator outputs 120/80 mmHg ART signals and 120/0 mmHg LV signals 
respectively to each IBP channel and check that the IBP wave is displayed correctly. 
8.  Repeat the preceding steps to test all IBP channels. 
 
 
Pressure Calibration 
Method 1: 
Tools required: 
  Medsim300B patient simulator, MPS450, or other equivalent equipment 
  Dedicated IBP adapter cable (300B, P/N 00-002199-00) (use P/N 00-002198-00, if the simulator is 
MPS450)