BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual          6-13
 
NOTE 
  Use care when disconnecting the cable to avoid damage. 
  Use care when removing the Wi-Fi Module from the Wi-Fi Support board to avoid 
damage to the connector. 
Provides application tips or other useful information to ensure 
that you maintain your product better. 
 
 
6.5.5 Removing the Internal Module Rack Assembly 
NOTE 
  Release the cable clips on the socket before attempting to disconnect the cable. Failure to do 
so will damage the cable. 
 
  N12 
1.  Disconnect the battery interface cable from the main board, then disconnect the internal module 
rack COMM board cable. Remove four M3x6 screws and lift up to remove the battery holder 
assembly. 
 
2.  Loosen and remove the two M3X6 screws, and remove the battery interface board.   
 
Cable of battery interface board 
J10 
J20 
Cable of internal module 
rack COMM board