BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual          3-15 
6.  Select the Calibrate button next to the desired IBP channel to start a calibration. 
If the calibration is completed successfully, the message Calibration Completed! will be displayed. 
Otherwise, a corresponding message will be displayed. 
 
 
3.3.7 C.O. Test 
Tools required: 
  Medsim300B patient simulator or other equivalent simulator 
  C.O. adapter box (for 300B) 
 
1.  Connect the patient simulator to the C.O. module using a C.O. main cable. 
2.  Set the blood temperature (BT) to 37ºC on the patient simulator and check the temperature value is 
37±0.1ºC. 
3.  Switch off Auto TI and adjust TI (IT) to 24ºC. Select C.O. Measure to enter the C.O. measurement 
window and set Comp. Const. to 0.595.   
4.  Set the injectate temperature to 24ºC and the C.O. to 5 L/min on the C.O. simulator. Select Start in 
the C.O. measurement window to start C.O. measurements, and press the run key on the simulator 
after 3-10 seconds. 
6.  Verify that the C.O. value is 5±0.25 L/min. 
 
 
3.3.8 Mainstream CO
2
 Tests 
 
NOTE 
  Make sure that the barometric pressure set in Main Menu 
  Maintenance 
  enter the 
required password 
  Other accords with the local barometric pressure before performing 
mainstream CO
2
 tests. 
 
Tools required: 
  A gas cylinder with 5±0.03% CO2, 21.0% O2 and balance gas N2 (P/N 0075-00-0033-01), or a steel gas 
cylin der wit h :  
  CO2 concentration 3% - 7%    
  a/c  ≤  0.01 (wh e re  a  =  ab so lu t e  ga s co nce n t ra t io n  a ccu ra cy, c =  g as co n ce n tra t io n )    
  balance gas N2 
  T-shape connector 
  Tubing 
  Flowmeter 
 
1.  Plug the module into the module rack and connect the sensor.