BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual          4-7
4.6.6 Power Supply Failures 
Failure Description  Possible Cause  Troubleshooting 
Battery cannot supply 
power 
 
Battery damaged  Replace the battery. 
Cable defective or 
improperly connected
 
 
Verify the cable connecting the main control 
board to the battery interface board is 
correctly connected.
 
 
Verify the connecting cables and connectors 
are not damaged.
 
Battery cannot be 
recharged or cannot be 
fully recharged
 
Battery damaged  Battery change 
Cable defective or 
improperly connected
 
  Verify the cable connecting the main control 
board to the battery interface board is 
correctly connected.
 
 
Verify the connecting cables and connectors 
are not damaged.
 
Main control board 
failure
 
Replace the main control board. 
 
 
 
4.6.7 Network Related Problems 
Failure Description  Possible Cause  Troubleshooting 
The patient monitor 
cannot be connected 
to iView system
 
   
   
Frequent dropouts 
Improper network cable 
connection
 
Check for network cable connection and 
length (which should not exceed 50 m), or 
check whether the laid network cable is too 
near to the power supply for large power 
equipment.
 
Incorrect network settings 
Check for IP conflict in the network. If conflict 
is found, reset the network.
 
The patient monitor is 
connected to a 
network but cannot 
view other patients in 
the View Others mode
 
Improper network cable 
connection
 
Check for network cable connection and 
length (which should not exceed 50 m), or 
check whether the laid network cable is too 
near to the power supply for large power 
equipment.
 
Too many simultaneous 
requests for viewing the 
patient monitor
 
Confirm the maximum number of 
simultaneously connected monitors 
according to the user manual.
 
Incorrect network settings 
Check for IP conflict in the network. If conflict 
is found, reset the network.