BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual          3-23 
3.7 Recorder Check 
Tools required: 
  None. 
 
1.  Print ECG waveforms. The recorder should print correctly and the printout should be clear. 
2.  Set the recorder to some problems such as out of paper, etc. the patient monitor should give 
corresponding prompt messages. After the problem is removed, the recorder should be able to work 
correctly. 
3.  Switch automatic alarm recording for each parameter ON and then set each parameter's limit outside 
set alarm limits. Corresponding alarm recordings should be triggered when parameter alarms occur.   
 
 
3.8 Network Print Test 
NOTE 
  HP LaserJet Pro M202dw laser printer is recommended for BeneVision patient monitor series 
 
 
3.8.1 Device Connection and Setup 
1  Connect the patient monitor and network printer to a network switch using common network cables 
as follows: 
 
2  Select Main Menu  → Maintenance  →  enter the required password  →Network Setup and set the 
IP address of the patient monitor in the same network segment with that of the network printer. (See 
the instructions for use accompanying the printer) 
3  Select Main Menu  → Maintenance  →  enter the required password  →  Print and set the IP address 
of the printer to the actual IP address, and set the paper size to the actual size.   
4  Set the print resolution to 300dpi or 600dpi as required.   
5  Click Print Test Page to check whether the output of the printer's test page is normal. If not, recheck 
the connection and configuration of the printer.   
 
 
BeneVision 
monitor 
Network 
printer 
Switch 
Network cable 
Network cable