EasyManuals Logo

Mindray Datascope AS 3000 User Manual

Mindray Datascope AS 3000
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
Troubleshooting Chart Repair Information
3 - 6 0070-10-0683 AS3000™ Service Manual
“Inspiration Sensor . . . FAIL”
or
“Inspiration Sensor . . . FAIL
Try re-starting the system with
all gas flows OFF” (UI
version 2.25 and higher)
Startup test 1 Fresh gas is flowing during the startup
procedure.
2 The tubing kinked or occluded.
3 The sensor board is defective.
1 Verify that no fresh gas is flowing during the
startup procedure.
2 Check the tubing for kinks and remove any
occlusions.
3 Replace the sensor board and then perform
each item of the calibration and startup test
again.
“O
2
Supply . . . FAIL” Startup test
Runtime
1 The gas source is not connected.
2 The gas supply pressure switch or its cable
is damaged.
3 The amplifier board is defective.
1Check gas source.
2 Check and replace the switch or cable of the
gas source.
3 Replace the amplifier board, and then perform
each item of the calibration and startup test
again.
“AC Power Failure” Startup test
Runtime
1 AC power is not connected.
2 A fuse is tripped
1 Check the connection of AC power.
2 Replace the fuse.
“Low Battery” Runtime 1 Battery capacity is low.
2 The battery is defective.
3 The connection wire or power board is
defective.
1 Charge the battery by connecting AC power.
2 Replace the battery.
3 Check the connection wire or replace the
power board.
“Alarm Speaker . . . FAIL” Startup test 1 The connection wire to the speaker is
disconnected or damaged.
2 The speaker is damaged.
3 The keyboard board is defective.
1 Check whether the connection wire of the
speaker is disconnected or damaged.
2 Replace the user interface assembly.
3 Replace the user interface assembly.
“Software Mismatch”
or
“Software Mismatch or
Failure to Shut Down
Completely” (UI version 2.25
and higher)
Startup test 1 Power is recycled to the unit too quickly, not
allowing the memory to fully clear.
1 Shut down the unit and wait until the backlight
is fully extinguished before restoring power.
Charging state is always
displayed.
Runtime 1 Battery capacity is low.
2 The connection wire to the battery is
disconnected or damaged.
3 The battery is defective.
1 Charge the battery by connecting AC power.
2 Check whether the connection wire to the
battery is disconnected or damaged.
3 Replace the battery.
During startup leak test, the
bellows moves when it
shouldn’t
Startup test 1 A solenoid failure or wiring error. 1 Correct the wiring or replace the solenoid
valve.
In leakage detection, The APL
Valve detection pressure
displayed on the pressure
gauge does not stay within
28 - 32 cmH
2
O.
Startup test 1 The Breathing System pneumatic hose leaks
gas when in manual mode.
2 Pressure gauge failure.
3 The APL Valve is out of spec
1 Replace the Breathing System pneumatic hose
2 Replace the pressure gauge
3 Replace the APL Valve
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals