EasyManua.ls Logo

Mindray Datascope AS 3000

Mindray Datascope AS 3000
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AS3000™ Service Manual 0070-10-0683 5 - 7
Calibration Calibration Procedures
5.5.3 Flow Sensor Calibration
1. Preparing the unit
a. Remove the Inspiratory and Expiratory valve disks, and reinstall the valve covers.
b. Connect a Respiration Tube between the Inspiratory and Expiratory Ports. (see
FIGURE 5-2)
c. Use the AS3000 Test Plug to block the Bag Port. (see FIGURE 5-3)
d. Set the APL Valve to 70 cmH
2
O.
e. With the Fresh Gas flow set to 100 mL/min, press the O
2
Flush button. Verify that the
Paw gauge reads a minimum of 40 cmH
2
O after releasing the button.
f. If 40 cmH
2
O cannot be attained, open and close the APL Valve and repeat step e
prior to leakage troubleshooting.
g. Set the Fresh Gas flow to 0 mL/min.
h. Select Continue.
FIGURE 5-2 Connecting the
Respiration Tube
FIGURE 5-3 Blocking the Bag Port

Table of Contents

Related product manuals