Calibration Warnings, Precautions, and Notes Calibration
5 - 2 0070-10-0683 AS3000™ Service Manual
5.2 Calibration Warnings, Precautions, and Notes
5.2.1 Warnings
WARNING: For continued protection against fire hazard, replace all
fuses with the specified type and rating.
WARNING: In order to prevent an electric shock, the machine
(protection class I) may only be connected to a correctly
grounded mains connection (socket outlet with grounding
contact).
WARNING: Remove all accessory equipment from the shelf before
moving the anesthesia machine over bumps or on any
inclined surface. Heavy top loading can cause the machine
to tip over causing injury.
WARNING: Possible explosion hazard. Do not operate machine near
flammable anesthetic agents or other flammable
substances. Do not use flammable anesthetic agents (i.e.,
ether or cyclopropane.)
WARNING: The use of anti-static or electrically conductive respiration
tubes, when utilizing high frequency electric surgery
equipment, may cause burns and is therefore not
recommended in any application of this machine.
WARNING: Possible electric shock hazard. The machine may only be
opened by authorized service personnel.
WARNING: Compressed gasses are considered Dangerous Goods/
Hazardous Materials per I.A.T.A. and D.O.T. regulations. It is
a violation of federal and international law to offer any
package or over pack of dangerous goods for
transportation without the package being appropriately
identified, packed, marked, classified, labeled and
documented according to D.O.T. and I.A.T.A. regulations.
Please refer to the applicable I.A.T.A. Dangerous Goods
Regulations and /or the Code of Federal Regulations 49
(Transportation, Parts 171-180) for further information.
5.2.2 Cautions
CAUTION: Refer to the “Periodic Maintenance Schedule of Service
Activities” on page 6-2, in the Periodic Maintenance section
for assistance when performing scheduled periodic
maintenance.
CAUTION: Do not leave gas cylinder valves open if the pipeline supply
is in use and the system master switch is turned to 'ON'. If
used simultaneously, cylinder supplies could be depleted,
leaving an insufficient reserve supply in the event of
pipeline failure.
CAUTION: Use cleaning agent sparingly. Excess fluid could enter the
machine, causing damage.
CAUTION: This machine must only be operated by trained, skilled
medical staff.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals