EasyManua.ls Logo

Mindray Datascope AS 3000

Mindray Datascope AS 3000
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ventilator UI Theory of Operation
1 - 34 0070-10-0683 AS3000™ Service Manual
1.7.2 Display
The display is a 10.4 inch TFT LCD
5. ON/OFF Backlight ENABLE
6. DIM Contrast adjust
Power of 232 Isolator, J9
PIN NAME FUNCTION
1. 5V Power 5V
2. NC No connection
3. GND Ground
COM1/2
PIN NAME FUNCTION
2 RXD 232RXD TO PC104 INPUT
3 TXD 232TXD TO PC104 OUTPUT
5 GND Ground
Test Point Definition
DESIGNATOR NAME FUNCTION
T5 +5V Logic Power 5V
T6 P12V Power 12V
T9 PWBUS Power Bus
LED Designation
DESIGNATOR STATUS FUNCTION
LT1 GREEN/BLINK Keyboard ready
Backlight Power, J2 (Continued)
PIN NAME FUNCTION

Table of Contents

Related product manuals