Calibration Procedures Calibration
5 - 10 0070-10-0683 AS3000™ Service Manual
5.5.5 Paw Sensor
1. 0 cmH
2
O Calibration
FIGURE 5-6 Removing the 3L bag
a. Remove the 3L bag from the breathing circuit.
b. Set the APL Valve to 0 cmH
2
O.
c. Select Continue to start calibration.
d. Follow the on-screen instructions.