DICOM 10-5
DICOM storage setting items are described as follows:
Name
Description
Device
After you set the server (s) in DICOM Server Setting, the name
(s) will appear in the drop-down list, select the name of the
storage server.
Service Name Default is xxx-Storage, and it can be modified.
AE Title
Application Entity title, Here, it should be consistent with that of
the storage server.
Port
DICOM communication port, 104 by default. Here, the port should
be consistent with that of the storage server port.
Maximum Retries Range: 0-9.
Interval Time(s)
Reserved feature.
Timeout
Refers to time after which the system will stop trying to establish
a connection to the service. Value: 560s, in increments of 5s,
and 15 by default.
Color Mode
Gray/ Mixed/ Color
Compression Mode
Select the compression mode: uncompressed, RLE, JPEG and
JPEG2000.
Compression Ratio
Select the compression ratio: lossless, low, medium and high.
Allow Multiframe
If SCP supports this function, then select it.
Max Framerate 25, 30, 35, Full are available, user-changeable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-50 and is the answer not in the manual?

Mindray DP-50 Specifications

General IconGeneral
Transducer Ports2
Cine LoopYes
TypePortable Ultrasound System
ProbesConvex, Linear
ConnectivityUSB, DICOM
Power Supply100-240V~ 50/60Hz
ApplicationsObstetrics, Gynecology, Urology, Cardiology