Safety Precautions 1-7
The ultrasound system can be powered by connecting to
mains power supply or powered by internal battery, if you
have any doubt with stability of mains power protective
grounding, please use internal battery.
DO NOT use the system in the vicinity of strong electromagnetic field (such
as a transformer), which may affect the performance of the system.
DO NOT use the system in the vicinity of high-frequency radiation source,
which may affect the performance of the system or even lead to failure.
3.
When using or placing the system, keep the system horizontal to avoid
disbalance.
4. To avoid damaging the system, DO NOT use it in following environment:
(1) Locations exposed to direct sunlight;
(2) Locations subject to sudden changes in environmental temperature;
(3) Dusty locations;
(4) Locations subject to vibration;
(5) Locations near heat generators;
(6) Locations with high humidity.
Turn ON the system only after the power has been turned OFF for a while. If
the system is turned ON immediately after being turned OFF, the system may
not be rebooted properly and could malfunction.
6. Remove ultrasound gel from the face of a probe when the examination is
complete. Water in the gel may enter the acoustic lens and adversely affect
the performance and safety of the transducer.
7. You should properly back up the system to a secure external storage media,
including system configuration, settings and patient data. Data stored to the
system’s hard drive may be lost due to system failure, improper operation or
accident.
Do not apply external force to the control panel, otherwise, the system may
be damaged.
If the system is used in a small room, the room temperature may rise. Please
provide proper ventilation and free air exchange.
10.
To dispose of the system or any part, contact Mindray Customer Service
Department or sales representative. Mindray
is not responsible for any
system content or accessories that have been discarded improperly. Mindray
is not responsible for any system content or accessories that have been
discarded improperly.
Electrical and mechanical performance may be degraded due to long usage
(such as current leakage or distortion and abrasion), the image sensitivity and
precision may become worse too. To ensure optimal system operations, it is
recommended that you maintain the system under a Mindray service
agreement.
Ensure that the current exam date and time are the same as the system date
and time.
13. DO NOT turn OFF the power supply of the system during printing, file storage
or invoking other system operations. An interrupted process may not be
completed, and can become lost or corrupted.