A-4 Electrical Safety Inspection
Reversed Polarity.
LIMITS
For BF applied parts: 5000 μA.
A.9 Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the
remaining Applied Part connectors. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition);
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF
applied parts,
100μA in Normal Condition.
500μA in Single Fault Condition.
Note:
Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
Follow the instructions of the analyzer manufacturer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-50 and is the answer not in the manual?

Mindray DP-50 Specifications

General IconGeneral
Transducer Ports2
Cine LoopYes
TypePortable Ultrasound System
ProbesConvex, Linear
ConnectivityUSB, DICOM
Power Supply100-240V~ 50/60Hz
ApplicationsObstetrics, Gynecology, Urology, Cardiology

Related product manuals