Hardware Introduction 
2-15 
overcurrent/short-circuit protection works, the output stops and does not affect other outputs. 
They except HV can recover automatically after the failure is removed. 
2.3.5. Maintenance 
Before test or service, connect the power supply board as the figure below shows. The 
electronic load A is necessary while the electronic load B is not. When adding the electronic 
load B, ensure the load for 390VDC does not exceed 0.1A and is added only when loads for 
other outputs are minimum.   
 
You can add loads only for +5V, +13.5V, +12V and HV outputs. It is recommended to start 
debugging when the load is a little heavier than the minimum and to increase the load during 
the debug if necessary. 
After connecting the board properly, you can locate the failure and service the board 
according to the procedure shown in the figure below. The following procedure is just a basic 
instruction and you should operate according to the actual situation.