10
T1 Quick Reference Guide
c. Select Main Menu (quick key) or to return to the main display
Zero Invasive Blood Pressure
To zero an invasive blood pressure transducer on the T1 monitor:
a. Connect the invasive blood pressure cable to the invasive blood pressure
transducer set
b. Position transducer and open the invasive line in accordance with
hospital policy
c. Select displayed invasive blood pressure waveform or IBP numeric tile to
enter the invasive blood pressure Setup menu
d. Scroll and select Zero >> in the specific invasive blood pressure setup
menu
e. Select Zero
f. Select Main Menu
(quick key) or to return to the main display
ALARMS
Adjust Alarm Limits
The Alarm Setup tab allows the clinician to configure alarm limits and alarm
responses on the T1 monitor.
To adjust a patient’s alarm limit on the T1 monitor:
a. Select parameter’s waveform or numeric tile to open parameters Setup
menu
b. Select Alarm Setup >>
c. Select High or Low limit value and adjust using the up and down arrows
d. Select Main Menu
(quick key) or to return to the main display
Or
a. Scroll and select Alarm Setup >> from Main Menu
(quick key)
b. Select Parameters tab to change parameters or ST Alarm tab to change
ST
c. Scroll and select the parameter
d. Select High or Low limit value and adjust using the up and down arrows
e. Select Main Menu
(quick key) or to return to the main display