20
T1 Quick Reference Guide
CLEANING/DISINFECTING
Cleaning the Main Unit/Module/Handle/Docking Station
Clean your equipment on a regular basis. Before cleaning, consult your
hospital's regulations for cleaning the equpment.
The following table lists approved cleaning and disinfecting agents.
Product Name Product Type Ingredients
Sodium hypochlorite
bleach*
Liquid Sodium hypochlorite bleach
0.5%
Hydrogen peroxide* Hydrogen peroxide 3%
Isopropanol Isopropanol 70%
1-Propanol* 1-Propanol 50%
CIDEX OPA Solution Ortho-Phthalaldehyde 0.55%
Metrex CaviCide1™ Diisobutylphenoxyethoxyethyl
dimethyl benzyl ammonium
chloride 0.28%,
Isopropanol 17.2%
Virex® II 256+ Didecyl dimethyl ammonium
chloride 8.704%,
n-Alkydimethyl benzyl
ammonium chloride 8.190%
Virex®TB n-Alkyl dimethyl benzyl
ammonium chlorides 0.105%,
n-Alkyl dimethyl ethylbenzyl
ammonium chlorides 0.105%
Rely+On™ Virkon®Powder
*(Used as 1% solution)
Powder Used as 1% solution
Biocidal active:
Pentapotassium bis
(peroxymonosulphate) bis
(sulphate)(500g/kg),
Contains dipotassium
peroxodisulphate.