T1 Quick Reference Guide
Adjust Arrhythmia Alarms (Arrhythmia Option Required)
To adjust a patient’s arrhythmia alarm on the T1 monitor:
a. Scroll and select Alarm Setup >> from Main Menu
(quick key)
b. Select Arrh. Analysis tab
c. Scroll and select individual arrhythmia
d. Select
•Alarm – to enable or disable arrhythmia alarm responses
•Alm Lev – to adjust the audio and visual alarm indicators. Level options
include High (Red), Medium (Yellow), Low (Yellow), and Message
e. Select Main Menu
(quick key) or to return to the main display
To adjust a patient arrhythmia threshold setting on the T1 monitor:
a. Scroll and select Alarm Setup >> from Main Menu
(quick key)
b. Scroll and select Arrh. Threshold tab
c. Scroll and select individual arrhythmia threshold
d. Adjust value using the up and down arrows
e. Select Main Menu
(quick key) or to return to the main display
DPM Central Station Interface: Adjusting arrhythmia alarms on the DPM
Central Station or WorkStation will not change arrhythmia alarms on the T1
bedside monitor. Adjusting arrhythmia alarms on the T1 bedside monitor will
affect the arrhythmia alarms on the DPM Central Station.
Panorama Central Station Interface: Adjusting arrhythmia alarms on the
Panorama Central Station or WorkStation will not change arrhythmia alarms
on the T1 bedside monitor. Adjusting arrhythmia alarms on the T1 bedside
monitor will affect the arrhythmia alarms on the Panorama Central Station and
WorkStation.
Disable Non-Lethal Arrhythmias (Arrhythmia Option Required)
To disable all arrhythmia alarms except lethal arrhythmias on the T1 monitor:
a. Scroll and select Alarm Setup >> from Main Menu
(quick key)
b. Scroll and select Arrh. Analysis tab
c. Select Lethals Only
d. Select Main Menu
(quick key) or to return to the main display