VS 8/VS 8A Vital Signs Monitor Operator’s Manual A - 1
A Product Specifications
A.1 Monitor Safety Specifications
The equipment is classified, according to IEC60601-1:
A.2 Environmental Specifications
• The equipment may not meet the performance specifications if stored or
used outside the specified temperature and humidity ranges. If the
performance of the equipment is degraded due to aging or environmental
conditions, contact your service personnel.
• When the monitor and related products have differing environmental
specifications, the effective range for the combined products is that range
which is common to the specifications for all products.
Type of protection against
electrical shock
CLASS I EQUIPMENT, equipment energized from an
external and internal electrical power source.
Degree of protection against
electrical shock
DEFIBRILLATION-PROOF TYPE CF AAPPLIED PART for
SpO
2
, NIBP, and SmarTemp™ Temp
DEFIBRILLATION-PROOF TYPE BF APPLIED PART for
Exergen TemporalScanner ™ Temp.
Mode of operation Continuous
Degree of protection against
harmful ingress of water
IPX2 (Protection against vertically falling water drops
when ENCLOSURE tilted up to 15°)
Degree of safety of application
in the presence of a
FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTUREWITH AIR or WITH
OXYGEN OR NITROUS OXIDE
Degree of mobility Portable
• The environmental specifications of unspecified parameter modules are the
same as those of the main unit.
Main Unit
SmarTemp™
Temperature Module
Exergen TemporalScanner Thermometer
A.3 Power Supply Specifications
AC Power
Item Operating conditions Storage conditions
Temperature 0 to 40 ºC –20 to 60 ºC
Relative humidity
(non-condensing)
15% to 95% 10% to 95%
Barometric 427.5 mmHg to 805.5 mmHg (57.0
kPa to 107.4 kPa)
120 to 805.5 mmHg (16.0 kPa
to 107.4 kPa)
Item Operating conditions Storage conditions
Temperature 5 ºC to 40 ºC Same as the main unit
Relative humidity (non-condensing) Same as the main unit Same as the main unit
Barometric Same as the main unit Same as the main unit
Item Operating conditions Storage conditions
Temperature 15.5ºC to 40 ºC (60°F to
104°F)
–20ºC to 50 ºC (-4°F to
122°F)
Relative humidity (non-condensing) up to 95% up to 95%
Line voltage 100 to 240 VAC (±10%)
Current 0.9 to 0.5 A
Frequency 50/60 Hz (±3 Hz)
Fuse T2AL-250V