VS 8/VS 8A Vital Signs Monitor Operator’s Manual 18 - 7
The following table lists approved Masimo SpO
2
cable cleaning and disinfecting agents:
The following table lists approved disinfectants for the Nellcor SpO
2
cables:
PDI Super Sani-
Cloth® Germicidal
Disposable Wipe
Wipes n-Alkyl dimethyl ethylbenzyl ammonium chlorides
0.25%,
n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,
Isopropyl Alcohol 55.0%
Product Name Product Type Active Ingredients
Isopropanol Liquid Isopropanol 70%
Product Name
Product
Type
Active Ingredients
Sodium hypochlorite bleach Liquid Sodium hypochlorite bleach, 0.5%
Isopropanol Liquid Isopropanol 70%
1-Propanol Liquid 1-Propanol 50%
Virex® TB Liquid
n-Alkyl dimethyl benzyl ammonium
chlorides 0.105%,
n-Alkyl dimethyl ethylbenzyl ammonium
chlorides 0.105%
Rely+On™ Virkon® High
Level Surface Disinfectant
Powder
Used as 1% solution*
Biocidal active:
Pentapotassium bis
(peroxymonosulphate) bis (sulphate)
(500g/kg)
Contains dipotassium peroxodisulphate.
Clorox Dispatch® Hospital
Cleaner Disinfectant Towels
with Bleach
Wipes
Sodium Hypochlorite 0.65%
Clorox Healthcare® Bleach
Germicidal Wipes
Wipes
Sodium Hypochlorite 0.55%
Clorox Healthcar
Hydrogen Peroxide Cleaner
Disinfectant Wipes
Wipes
Hydrogen Peroxide 1.4%
Product Name
Product
Type
Active Ingredients
18 - 8 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
18.6.2 Cleaning the Accessories
You should clean the accessories (NIBP air hose, and SpO
2
cable) on a regular basis.
Before cleaning the accessories, consult your hospital’s regulations for cleaning the
accessories.
To clean the accessories (NIBP air hose and SpO
2
cable), follow this procedure:
1. Clean the accessories with wipes or a soft cloth moistened with one of the
cleaning agents listed in 18.3.1 Approved Cleaning and Disinfecting Agents only.
2. Wipe off all the cleaning agent residue with a dry cloth.
3. Allow the accessories to air dry.
18.6.3 Disinfecting the Accessories
We recommend that the accessories (NIBP air hose and SpO
2
cable) should be
disinfected only when necessary as determined by your hospital’s policy, to avoid long
term damage to the accessories. Cleaning the accessories before disinfecting is
recommended.
18.7 Sterilization
Sterilization is not recommended for this monitor, related products, accessories, or
supplies unless otherwise indicated in the Instructions for Use that accompany the
products, accessories or supplies.
Diversey Oxivir® TB Wipes
Wipes
Hydrogen Peroxide 0.5%
PDI Super Sani-Cloth®
Germicidal Disposable Wipe
Wipes
n-Alkyl dimethyl ethylbenzyl ammonium
chlorides 0.25%,
n-Alkyl dimethyl benzyl ammonium
chlorides 0.25%,
Isopropyl Alcohol 55.0%
Product Name
Product
Type
Active Ingredients

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray VS 8 and is the answer not in the manual?

Mindray VS 8 Specifications

General IconGeneral
BrandMindray
ModelVS 8
CategoryMonitor
LanguageEnglish

Related product manuals