VS 8/VS 8A Vital Signs Monitor Operator’s Manual 16 - 5
◆ Local Password: the monitor’s password for accessing the Maintenance
dialog is required.
◆ Clinician Authentication: the user name and password saved in the MLDAP
server are required.
You can select Modify Local Password to change the monitor’s password for accessing
the Maintenance dialog.
16.3.3 Selecting Password for Changing Alarm settings
You can select what password is used when changing alarm settings, including setting
alarm switch, alarm limit, and alarm priority. To do so, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the Authorization Setup tab → Authorization Setup tab.
3. From the Clinic area, set Alarm Setup to select the password for changing alarm
settings.
◆ No Password: changing alarm settings is not password protected.
◆ Local Password: The monitor’s password is required for changing alarm
settings.
◆ Clinician Authentication: The account and password in the MLDAP are
required for changing alarm settings.
If Alarm Setup is password protected, you can set Retention Time to select the timeout
period of password for accessing the alarm settings.
If there is no operation after the
specified timeout period is reached, you need to re-enter the password.
16.3.4 Selecting Password for Viewing Discharged Patient Data
You can select what password is used when viewing discharged data. To do so, follow
this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the Authorization Setup tab → Authorization Setup tab.
3. From the Clinic area, set View Discharged Patients to select the password for
viewing discharged patient data.
◆ No Password: no password is required for viewing discharged patient data.
◆ Clinician Login: discharged patient data can be viewed only after clinician
login succeeds.
16 - 6 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
16.4 Changing Settings about Saving and Sending Data
16.4.1 Setting Items about Saving Data
16.4.1.1 Setting Confirmation before Saving Physiological Data
You can set the monitor to require confirmation before saving physiological data. To do
so, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the Save&Send tab.
3. Switch on Confirm Before Saving.
16.4.1.2 Setting Clinician and Patient Information as Required
You can set clinician and patient information as required for saving physiological data.
To do so, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the Save&Send tab.
3. Switch on Clinician And Patient info. Required.
16.4.1.3 Enabling Automatic Sending after Saving Data
You can set the monitor to automatically send the data to preset destination after saving data.
To do so, follow this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the Save&Send tab.
3. Switch on Send After Saving.
16.4.1.4 Enabling Automatic Recording after Saving Data
You can set the monitor to automatically record the data after saving data. To do so, follow
this procedure:
1. Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2. Select the Save&Send tab.
3. Switch on Record After Saving.
16.4.1.5 Enabling Automatic Logout after Saving Data
You can set the monitor to automatically log out of the clinician account after saving data. To
do so, follow this procedure: