VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 1
C Default Settings
C.1 Parameters Default Settings
C.1.1 SpO2 Default Settings
Item Default Setting
Alarm SpO2 On/Off On
High Adult and pediatric: 100%
Neonate: 95%
Low 90%
Priority Med
SpO2 Desat On/Off On
Low 80%
Priority High
NIBP Simul Off
Sat-Seconds (for Nellcor SpO
2
)Off
Setup Sensitivity (for Masimo SpO
2
)APOD
Display PI (for Masimo SpO
2
)On
Speed 25 mm/sec
Averaging (for Masimo SpO
2
)8 s
Display SIQ (for Masimo SpO
2
)Off
Fast SAT (for Masimo SpO
2
)Off
C - 2 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
C.1.2 PR Default Settings
C.1.3 Temp Default Settings (SmarTemp)
C.1.4 NIBP Default Settings
Item Default Setting
PR Alarm PR On/Off On
High Adult: 120
Pediatric: 160
Neonate: 200
Low Adult: 50
Pediatric: 75
Neonate: 100
Priority Med
PR Setup
Pulse Volume 2
Item Default Setting
Alarm Temp On/Off Off
High 38.0 °C
Low 35.0 °C
Priority Med
Item Default Setting
Alarm NIBP-S On/Off On
High Adult: 160 mmHg
Pediatric: 120 mmHg
Neonate: 90 mmHg
Low Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 40 mmHg
Priority Med