VS 8/VS 8A Vital Signs Monitor Operator’s Manual C - 3
Alarm
NIBP-D On/Off On
High Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
Low Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority Med
NIBP-M On/Off On
High Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
Low Adult: 60 mmHg
Pediatric: 50 mmHg
Neonate: 25 mmHg
Priority Med
NIBP-S Extreme On/Off Off
High Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 35 mmHg
Priority High
NIBP-D Extreme On/Off Off
High Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low Adult: 35 mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority High
Item Default Setting
C - 4 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
Alarm NIBP-M
Extreme
On/Off Off
High Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low Adult: 45 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority High
Setup Initial Pressure Adult: 160 mmHg
Pediatric: 140 mmHg
Neonate: 90 mmHg
Interval
Sequence
Start Mode Clock
NIBP End Tone Off
Venipuncture Pressure Auto
Display Format Sys/Dia(Mean)
Display Alarm Limits Off
Patient Position Lying
Measurement Site Unspecified
Sequence Phase A
Duration
1 hr
Interval
5 min
Phase B
Duration
4 hrs
Interval
15 min
Phase C
Duration
4 hrs
Interval
30 min
Phase D
Duration
Continuous
Interval
1 hr
Phase E
Duration
Off
Interval
1 hr
Item Default Setting